Single and Dual Channel Moisture Measurement - Promet EExd

Promet EExd is certified as flameproof for use in IEC Zone 1 and 2/NEC Class 1 Div1. Promet EExd offers single and dual channel moisture measurement, with integrated electronic pressure reading, within a single field-installed flameproof housing. User interface is provided through the bright vacuum fluorescent display and touch screen controls

• EExd Flameproof ATEX and CSA certified

• Complete Zone 1 or 2 installation

• Low wattage AC supply only utility requirement

• Single- or dual-channel measurement

• Calibration traceable to NPL (UK) and NIST (USA)

• Field adjustment maintenance by user adjustment or unique sensor exchange service

• Range -120 to +30°C dew point, 0.001 to 30,000 ppmV

• Moisture content units in ppmV, mg/Nm-³, Lb/mmscf

• Integrated electronic pressure measurement

• Modbus RTU and 4-20 mA linear signals

• Process and instrument status volt-free alarm contacts

• Remote global access options

• Touch screen control

Natural Gas & Petrochemicals

Technical Details
Main Unit
Channel Configurations Single or two channel
Moisture Parameters Dew point °C & °F
Moisture Content Automatic pressure compensated conversions:
ppmv for natural gas and ideal gas
Lbmmscf and mg-m-3 for natural gas
Dew point at a pressure input by the
user for natural gas and ideal gas
Calculations for natural gas moisture content
based on either IS0 18453 or IGT#8 to
customer order preference
Analysis Pressure Barg, MPa and psig
Sensor Technology Michell Ceramic Moisture Sensor
Measurement Range -120ºC dp to +30ºC dp.
1 ppbV to 30,000 ppmv
0 to 250 Barg
Calibration Range -100ºC dp to +20ºC dp.
10 ppbV to 23,000 ppmv
Accuracy Dew point: ±1°C between –59.9 & +20°C dp
Moisture content: ±10 % of reading
Dew point: ±2°C between –60 & -100°C dp
Moisture content: ±20 % of reading
Analysis Pressure: ±0.25 % FS
Measured Resolution 0.1°C: –80 to +20°C dp
1°C: –100 to -80°C dp
Displayed Resolution Dew point: 0.1°C
Moisture content: autoscale, 5 digits
MPa and Barg: 0.1, psig: 1.0
Temperature coefficient Algorithm compensation
HMI
Keyboard Interface Display Capacitive touch-screen through glass window
Vacuum flourescent
Data Logging A rolling maximum of 150 data logs is available.
Each log records time, date, moisture and
pressure values for each channel.
5 minutes minimum and 60 minutes maximum
logging intervals can be set by the user
Communications & Output Two non-isolated 4-20 mA per measurement
channel. 500 Ω maximum load. Range and
parameter settable by user.
Modbus RTU @ 9600 baud-rate
Alarms: two volt free contacts per channel; one
process value and one instrument status. Also
available via Modbus communications
Main Unit Enclosure Internal temperature control for condensation
protection and stable analysis conditions
Type Flameproof EExd
Construction Cast copper free aluminium
Finish Chromate primer,Polyester P9010 powder
coated (black). Meets BS3900
Operating Pressure Max. 206 Barg
Sample Gas Connections ¼” NPT (female)
Sample Flow Rate 0.5 to 5.0 litres per min.
Sample flow alarm Optional
Power Supply 90 – 264Vac 47 to 63Hz, 110 to 300
Vdc, 180W
Operating environment Indoor/Outdoor -20°C to +60°C
Max 95 % rh. Shaded location
Hazadous Area Certification
ATEX Certification II 2 G
Code EEx d IIB + H2 T4 (135°C) IP66 Tamb -40 to +60°C
Certificate number TRL06ATEX21093X
cCSAus approved Div 1, Class 1, Group B,C,D
Certificate number 1722693
Remote Interface
Optional, for Indoor Safe Up to 31 Michell EExd Process
Area Installation Host
Instrumentation Range analyzers in any
capacity combination of models via Modcus
Optional, for Indoor Safe Area Installation Host Capacity Up to 31 Michell Process Instrumentation Range analyzers in any combination of models via Modcus
Display Backlit color 15 cm/6” LCD
User Interface Integral mouse, on-screen keyboard
Function View and modify analyzer operating settings.
Record parameters to file and view virtual
chart recorder Analyzer diagnostics
Data Download CSV (comma separated variable) format for
easy import to ExcelTM
Remote Access OPC (OLE for Process Control) over a LAN
Imbedded web page/unique IP address for
PC access
Power Supply 100-280V, 50/60Hz
Note: Specification are subject to change without notice.

Download Catalog PDF
Promet EExd

For all inquiries, you may contact us through Telephone, Email or Live Chat Support.

Operating Hours
Monday to Friday: 8.00am to 5.30pm

Technical Support (After Office Hour)
Tel: (65) 9876 6258

Contact Informations


By Phone / Fax By Email By Live Chat
Tel: (65) 6553 0366
Fax: (65) 6553 0828
This e-mail address is being protected from spambots. You need JavaScript enabled to view it



Site Search...
Select your Brand...
Select your Product...







Contact Info

Flotech Controls Pte Ltd
Process Control & Instrumentation

438 Tagore Industrial Avenue Singapore 787814 (
View Map)
Tel: (65) 6553 0366
Fax: (65) 6553 0828
Email: info@flotech.com.sg