Condumax II is a fully self-contained hydrocarbon dew-point analyzer, using the proven "Dark Spot" hydrocarbon dew-point measurement principle from the highly successful Condumax instrument. Combining a hydrocarbon dew-point cell and water dew-point measurement in the same compact and lightweight EExd enclosure, Condumax II requires nothing more than mains power to operate. Condumax II provides a continuous local indication of key measurements in addition to analogue and digital outputs for remote monitoring.
• Fully automatic on-line analysis
• Objective, repeatable measurements
• Fundamental cooled mirror principle
• Patented detection technique
• Self-cleaning
• CSA Certified
• ATEX approved to Ex II 2 G EExd IIB + H2 T4 for hazardous area operation
• Sampling system/sensor cell designed for exposed field location
• No purge air or external cooling system required
• Optional combined water and hydrocarbon dew-point analysis
• Modbus RTU communications protocol Touch screen interface
Natural Gas & Petrochemicals
| Technical Details | |
| Main Unit | |
| Hazardous Area Certification | |
| ATEX certified to: II 2 G EExd IIB + H2 T4 Tamb =60°C Certificate number TRL04ATEX11060X |
|
| CSA (US) certified to Class 1, Div 1, Groups B, C & D, T4. Certificate number 1722693 | |
| Hydrocarbon Dew-Point Measurement | |
| Measuring Technique | DARK SPOT™ fixed sample analysis. Direct photo-detection of hydrocarbon condensate at hydrocarbon dew-point temperature |
| Sensor Cooling | Automatic via 3-stage Peltier effect electronic cooler |
| Maximum Range | Up to Δ 55°C measurement depression from Main Unit operating temperature |
| Accuracy | ±0.5°C hydrocarbon dew point (Single condensible component analysis) |
| Sample Flow | 0.03m³/hr (0.5 Nl/min) - Alarm standard |
| Measurement Frequency | 6 cycles/hour (recommended) 12 cycles/hour (maximum) |
| Water Dew Point Measurement (Optional) | |
| Measuring Technique | Michell Ceramic Moisture Sensor |
| Range | Calibrated from -100 to +20°C dew point |
| Accuracy | ±1°C from -59 to +20°C dew point ±2°C from -100 to -60°C dew point |
| Sample Flow | 0.06 to 0.3 m3/hr (1 to 5 ln/min) - alarm standard |
| Pressure Measurement | |
| Units |
MPa, Barg, psig |
| Range | HCdp: 0 to 100 Barg Wdp: 0 to 200 barg |
| Accuracy | ±0.25 % FS |
| Hydrocarbon Dew-Point Analyzer | |
| Resolution | Hydrocarbon and water dew point: 0.1°C Pressure: 0.1 MPa and Barg, 1 psig |
| Sample Gas Supply | Natural gas up to 100 barg, pressure regulated in sampling system |
| Enclosure | EExd cast enclosure with removable glass window viewing port. Internally heated for condensation protection |
| Sample Gas Connections | ¼" NPT female ports for both hydrocarbon and water dew-point channels |
| Operating Environment | Indoors/Outdoors -20 to 60°C. Max 95% RH |
| Power Supply | 90-264 VAC 50/60 Hz, 200 W Main Unit; 300 W c/w indoor sampling system; 400 W c/w outdoor sampling system |
| Weight | Main Unit 22.5kg 42 kg (approx) c/w Sampling System (indoor) 57 kg (approx) c/w Sampling system (outdoor) |
| Integrated Display | Touch screen with vacuum fluorescent display |
| Outputs and Alarms | Modbus RTU, RS-485 @ 9600 baud rate Two 4-20 mA linear (non-isolated) outputs, user configurable for any combination of dewpoint or pressure parameters Hydrocarbon and water dew-point alarms via software register. Integrated low flow alarms for each sample flow |
| Remote Interface (Optional, for Indoor Safe Area Installation) | |
| Host Capacity | Up to 31 Michell EExd Process Instrumentation Range analysers in any combination of models via Modbus |
| Display | Backlit color 15 cm/6” LCD |
| User Interface | Integral mouse, on-screen keyboard |
| Function | View and modify analyser operating settings. Record parameters to file and view virtual chart recorder Analyser diagnostics |
| Ports | LAN, USB2 |
| Data Download | CSV (comma separated variable) format for easy import to Excel™ |
| Remote Access | OPC (OLE for Process Control) over a LAN Imbedded web page/unique IP address for PC access |
| Power Supply | 100-280 V, 50/60 Hz |
| Plant Control Integration | ActiveX components to be used by software programmers for integration of Modbus protocol into a general plant software system |
| Note: Specification are subject to change without notice. | |
For all inquiries, you may contact us through Telephone, Email or Live Chat Support.
Operating Hours
Monday to Friday: 8.00am to 5.30pm
Technical Support (After Office Hour)
Tel: (65) 9876 6258
Contact Informations
| By Phone / Fax | By Email | By Live Chat | |
| Tel: (65) 6553 0366 Fax: (65) 6553 0828 |
This e-mail address is being protected from spambots. You need JavaScript enabled to view it |
|